Measuring natural gas flow with a gas ultrasonic flowmeter (English Version)
Revision dimensions | Contents of the 2014 version | Changes in the 2023 version |
---|---|---|
Scope of application | Limited to specific places such as gathering and transportation devices | Added application scenarios such as new measurement standard devices |
Performance indicators | Does not distinguish between caliber differences | Specifies the different error limits for flow meters ≥300mm and <300mm |
Installation requirements | Ra value is not specified | New requirement for inner surface roughness of straight pipe section ≤3.2μm |
The flow velocity is calculated by measuring the time difference ΔT between the upstream and downstream sound pulses of the ultrasonic transducer and combining the formula V=L²ΔT/(2X·tU·tD).
Project | Cycle | Inspection content |
---|---|---|
Regular inspection | ≥1 time/year | Channel status, zero flow reading, transducer deposits |
Offset check | After each calibration | Deviation from the last calibration result (8.3.3) |
Expanded uncertainty of volume flow under standard reference conditions:
Ur(qn)=k×√[ur²(qf)+ur²(Pf)+ur²(Tf)+ur²(Zf)+ur²(Zn)+ur²(an)](k=2)
An is the additional installation uncertainty, and it is recommended to take 1/3 of the flow meter MPE (10.2.1)